| Abstract |
DOI: 10.1051/jcp:1999233
J. Chim. Phys. Vol. 96, N°9/10 p. 1559-1565
Investigation of the diffusion measurement using PFG and test of robustness against experimental conditions and parameters
S. Augé1, B. Amblard-Blondel2 and M.-A. Delsuc11 Centre de Biochimie Structurale, Faculté de Pharmacie, 15 avenue Charles Flahault, 34060 Montpellier Cedex 2, France
2 ELF Antar France, Chemin Canal, BP. 22, 69360 Solaize, France
Abstract
The DOSY experiment, based on pulsed-field gradient (PFG) NMR, is used to analyse the diffusion coefficients of species in
mixture. We present here an evaluation of this method. Various
experimental tests demonstrates that this method is robust against experimental parameters like relaxation delay, gradients
duration or pulse flip angle. Concerning the diffusion delay, it has to be
kept as short as possible because convection can appear during this delay. We show that the rotation of the sample permits
to minimize convection effect. Moreover, we propose various gradient
sampling strategies and show that they are equivalent for a model case.
Résumé
L'expérience RMN DOSY, basée sur les gradients de champs pulsés, est
utilisée pour l'analyse des coefficients de diffusion d'espèces en mélange.
Nous présentons ici une évaluation de cette méthode. Plusieurs tests
expérimentaux démontrent que cette méthode est robuste envers les
paramètres expérimentaux tels que le délai de relaxation, la durée des
gradients ou l'angle des impulsions. En ce qui concerne le délai de
diffusion, il doit être le plus court possible car durant ce délai de la
convection peut apparaître. Nous montrons que la rotation de l'échantillon
permet de minimiser l'effet de cette convection. D'autre part, nous
proposons plusieurs stratégies d'échantillonnage de l'intensité des gradients
et montrons que, dans un cas modèle, elles sont équivalentes.
Key words: DOSY, diffusion, pulsed field gradient, convection, method robustness.