Use of spontaneous Raman spectroscopy as a diagnostic of silica-particle-laden flames
Laboratoire de Chimie Physique de la Combustion, Université de Poitiers, URA 872 au CNRS, Domaine du Deffend, 86800 Mignaloux-Beauvoir, France
Spontaneous Raman Spectroscopy has been used as a diagnostic of flames containing high concentrations of silica particles. Because of the low emissivity of silica, no visible radiation from the particles has been observed, even at the highest temperature attainable by the burner (2050 K). So far, temperatures only have been measured from oxygen vibrational spectra. Spatially-resolved measurements of N2, O2, CO2 concentration and temperature are feasible, however, in (lames containing 107- 108 particles/cm3
Résumé
La Spectroscopie Raman Spontanée a été utilisée comme diagnostic de flammes contenant des concentrations élevées de particules de silice. En raison de la faible émissivité de la silice, aucune radiation due aux particules n'est observée dans le domaine visible, même à la température la plus élevée atteinte par le brûleur (2050 K). Jusqu'à maintenant, les températures seules ont été mesurées à partir des spectres de vibration de O2. Des mesures locales de concentration des espèces N2, O2, CO2, comme de température, sont toutefois réalisables dans des flammes contenant 107-108 particules/cm3